ARTIFACTS · LESSON 36

Artifact troubleshooting workflow.

Move from “something looks wrong” to one informed, testable correction without losing sight of safety or the clinical question.

About 18 minutesTRACE workflow4 clinical scenarios
TRACETARGET · RECOGNIZE · AXIS · CHANGE · EVALUATE
01EXPLAIN

DO NOT BEGIN BY RANDOMLY CHANGING PARAMETERS

First decide whether the problem is safety, anatomy, acquisition, reconstruction, or equipment.

An artifact can come from the patient, physiology, positioning, coil setup, sequence design, spatial encoding, reconstruction, hardware, or the surrounding RF environment. More than one cause may be present.

A strong troubleshooting process describes the pattern, compares it with the sequence and directions, forms a short cause list, changes one meaningful variable, and checks whether the diagnostic information improved.

SAFETY GATE

Patient safety comes before artifact correction.

If the image or situation raises a concern about an implant, foreign body, heating, pain, monitoring device, projectile risk, or other safety issue, pause the examination and follow site policy. Do not continue scanning simply to test an artifact theory.

IN PLAIN LANGUAGEDescribe first. Decide second. Change one thing. Check the result.
02VISUALIZE

THE TRACE WORKFLOW

Use the same five questions every time.

TTarget

What anatomy and diagnostic question must remain visible? Is there any reason to stop for safety?

RRecognize

Is the pattern a ghost, fold, bright-dark edge, signal void, narrow line, broad shade, ringing, distortion, or unusual noise?

AAxis and appearance

Which direction is phase and which is frequency? Does the artifact follow anatomy, an edge, a slice, every image, or only one sequence?

CChange one cause

Select the safest approved change most likely to test the leading explanation. Avoid changing several unrelated controls at once.

EEvaluate and escalate

Compare the result with the original, confirm diagnostic quality, document when required, and seek help if the cause persists.

03CONNECT

TURN APPEARANCE INTO A SHORT CAUSE LIST

Let direction, location, and repetition narrow the possibilities.

WHAT YOU SEEGhosts or smearing

Check phase direction, patient motion, respiration, pulsation, flow, timing, and suppression.

WHAT YOU SEEFolded anatomy

Check encoded FOV, phase or partition direction, oversampling, positioning, and coil coverage.

WHAT YOU SEEBright-dark fat edge

Check frequency direction, receiver bandwidth, field strength, fat suppression, and TE.

WHAT YOU SEEVoid or distortion

Check air-tissue interfaces, verified metal, sequence family, TE, bandwidth, voxel size, and shimming.

WHAT YOU SEELine or broad shading

Check the RF door and environment, coil connection and selection, centering, calibration, and system performance.

WHAT YOU SEEResidual aliasing or patchy noise

Check acceleration factor, active coil elements, calibration, patient movement, FOV, and reconstruction limits.

COMPARE BEFORE AND AFTER

When safe and appropriate, preserve the original image and note the single change used for the repeat. A controlled comparison teaches more than several simultaneous changes and makes persistent problems easier to communicate.

04REMEMBER

THE DETECTIVE, NOT THE GUESSER

Every artifact leaves evidence.

A detective does not rearrange the entire scene. They describe what happened, identify where and when it happened, test the strongest explanation, and compare the new evidence.

PATTERNDIRECTIONCAUSEONE CHANGERESULT

Remember: Change one thing for one reason, then prove whether it helped.

05APPLY

CLINICAL REASONING PRACTICE

Walk through TRACE before repeating the sequence.

Respiratory ghosts cross the liver in the phase direction. What is a focused first response?

Confirm the phase direction and breath-hold performance, then choose one approved motion-control strategy, such as coaching, timing, respiratory compensation, or a shorter acquisition, while preserving the diagnostic requirement.

A bright zipper line appears on multiple patients and sequences. What does that pattern suggest?

A recurring RF, hardware, or environmental source is more likely than patient anatomy. Check approved immediate causes such as the RF door and unapproved electronics, then document and escalate through the quality-control or service process if it persists.

An accelerated series shows residual wrap after the patient shifted in the coil. What should you investigate?

Check whether calibration still matches the patient and coil position, along with active coil elements, FOV, direction, and acceleration factor.

The artifact improves after one parameter change, but the anatomy becomes too blurry. Is the problem solved?

No. A successful correction must preserve the clinical information. Rebalance the protocol or seek guidance rather than accepting a technically cleaner but nondiagnostic image.

LESSON 36 · CHAPTER 8 COMPLETE

You can recognize common artifacts, test their likely causes, and protect the diagnostic purpose of the exam.

Educational references

Troubleshooting steps vary by scanner, coil, sequence, anatomy, patient condition, and facility policy. Use approved protocols, vendor guidance, quality-control procedures, and the appropriate clinical or service escalation path.